- Device Size Dimensions: 170 — 196 18 mm 290 ,196 ‚134 mm (with humidifier) Weight: < 1.5 kg < 2.5 kg (with humidifier)
- Product Use, Transport and Storage
- Operation
- Transport and Storage
- Temperature:
- 5°C to 35°C (41°F to 95°F)
- -25°C to 70°C (-13°F to 158°F)
- Humidity:
- 15% to 93% Non-condensing
- 15% to 93% Non-condensing
- Atmospheric Pressure:
- 760 to 1060 hPa
- 760 to 1060 hPa
- Mode of Operation Continuous Mode Comparison
- Work Mode
- Pressure
- LCD Screen
- U-25T
- CPAP, S, T, S/T
- 4-25 cm H2O
- 2.4 inches
- SD Card
- With a capacity 2G, the SD card can record patient data and fault information.
- Furthermore, the language pack stored on the SD card enables you to change the language of the device.
- AC Power Consumption
- 100 - 240 V AC, 50/60 Hz, 2.0 A max Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Shock Type BF Applied Part Degree of Protection Against Ingress of Water IP22 Pressure Range
- 4 to 30 hPa (in 0.5 hPa increments), 30 hPa for CPAP mode, 30 hPa for the rest modes under single fault conditions.
- Pressure Display Accuracy ± (0.5 hPa + 4%) Ramp The ramp time ranges from 0 to 60 minutes
- Sound Pressure Level <30 dB, when the device is working at the pressure of 10 hPa.
- Sound Power Level <38 dB,
- when the device is working at the pressure of 10 hPa.
- SpO2 Range: 0-100%
- The margin of error for SpO2 between 70% and 100% is ±3%.
- No strict accuracy requirements
- for SpO2 below 70%.
- Pulse Rate Range: 40-240 bpm
- Margin of Error: ± 3% Wavelengths
- Red: 663 nanometers
- Infrared: 890 nanometers
- Maximal Optical Output Power Less than 1.5 mw maximum average
- Tube Length: 6 ft. (1.83m)
- Heater Settings 1 to 5(95 to 167°F/ 35 to 75°C)
- Maximum Operating Pressure 30 hPa Pressure Drop with Humidifier <0.4 hPa at 60 LPM flow Humidifier
Performance
Humidity Output: No less than 10 mg H2O/L
Environmental Conditions: Maximum airflow, 35°C, 15% relative humidity Maximum Delivered Gas Temperature <43°C The Form and the dimensions of the Patient Connection Port The 22 mm conical air outlet complies with ISO 5356-1